Technical Data
· Fully computerized system
· Full touch screen software
· RS 232 integrated opto-isolated connection for maintenance
Blood flow control: 20 ml/min steps
Single pump 0 to 800 ml/min
Double pump 0 to 600 ml/min
Blood pressure:
Venous -400 to 880 mm Hg
Arterial -400 to 880 mm Hg
TMP -60 to 520 mm Hg
Heparin pump:
Syringe size 20, 30 & 50 ml
Flow rate 0.1 to 10 ml/hr
Alarm systems:
Air bubbles on venous line
Technique Ultrasound
Sensitivity bubbles ≥ 10 µl
Blood on arterial & venous lines:
Technique IR light
Blood leaks IR light
Dialysis Fluids:
Flow rate 300 to 800 ml/min
Temperature 35°C to 39°C
Ultrafilter MF-100-DC
Conductivity:
Total 12.1 to 15.8 mS/cm
First step 2.4 to 3.6 mS/cm
4.0 to 6.0 mS/cm
Ultra filtration control:
Absolute Single Pass Coriolis Force
UF rate 0.1 to 2.5 kg/hr
Accuracy ± 1 g/min or ± 1%
Dimensions & requirements:
Size HxWxD 1740 x 500 x 755 mm
Weight 98 kgs
Water quality > AAMI norms
Water Temperature °C to 39
Filter MF 900
Power requirements:
Voltage 115–230 VAC ± 10%
Frequency 50 to 60 Hz